Aplikacija:
Standard
Avtomatsko čiščenje:
Ne
Delovna temperatura:
0 - 60 °C
Diameter:
40 mm
Dolžina:
Inline sensor (TriClamp): 332 mm
Dolžina kabla:
10 m (optional extension cables available)
Garancija:
24 mesecev
Konfiguracija montaže:
TriClamp
Maksimalna temperatura:
60 °C
Material:
Optics Carrier and Sleeve: stainless steel 1.4460 / stainless steel 1.4571
Merilna metoda:
Combined multiple beam alternating light method with infrared diode system and beam focusing
Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm
Solids (TS): 90º and 120º scattered light measurement, wavelength = 860 nm
Merilno območje:
0,001 - 500 g/L
Turbidity (TRB): 0.001 - 9999 NTU
Solids (TSS): 0.001 - 500 g/L
Model:
TSS sc
Način montaže:
Weld On TriClamp Fitting
Natančnost:
Turbidity up to 1000 NTU: < 5 % of measurement value or ± 0.01 NTU, whichever is greater
NTU: 5 % of measured value or ± 0.01 NTU, whichever is greater
Območje tlaka:
< 10 bar or < 100 m
Odzivni čas T90:
1 - 300 s adjustable
Parameter:
Turbidity, Suspended Solids
Ponovljivost:
TSS content: < 4 %
Turbidity: < 3 %
Posebna značilnost:
With Triclamp
Posebni zapiski:
Installation note:
Distance sensor–wall
> 50 cm (Turbidity)
> 10 cm (TSS)
Pretok:
Max. 3 m/s (the presence of air bubbles affects the measurement)
Servisni interval:
1 h/mesec
Temperatura okolja:
0 - 60 °C (briefly 80 °C)
Teža:
Approx. 1.6 kg
Umeritev:
Turbidity (TRB): Factory calibrated
Solids (TS): To be calibrated by customer on site
Zero point: Permanently calibrated in the factory
Umertivena metoda:
Turbidity: Formazin or Stablcal Standard (at 800 NTU). Requires a calibration kit.
Suspended Solids: Sample specific, based on gravimetric analysis with a correction factor procedure.
Vsebina paketa:
Turbidity & Suspended Solids sensor, user manual
Združljivost kontrolnih enot:
SC200, SC1000, SC4500. All controllers sold separately