LDO sc Process Sensor for Dissolved Oxygen

LDO sc Process Sensor for Dissolved Oxygen
Številka izdelka: LXV416.99.20001
EUR Cena: Kontakt
Na voljo

Extremely reliable - 36-month warranty on probe

The optimised temperature sensor and the new 3D factory calibration make O2 measurement even more accurate

No electrolyte or membrane changes necessary

Remote functions for simple and convenient data transfer via Internet and SMS

Specifikacije

Delovna temperatura: 0 - 50 °C
Diameter: 48.25 mm
Dimenzija (d x g): 48.25 mm x 254 mm
Dolžina: 254 mm
Dolžina kabla: 10 m (33 ft)
Garancija: Probe body: 3 years
Sensor cap: 2 years
In standard applications
Interference : No Interferences from the following: H2S, pH, K+, Na+, Mg2+, Ca2+, NH4+, Al3+, Pb2+, Cd2+, Zn2+, Cr (total), Fe2+, Fe3+, Mn2+, Cu2+, Ni2+, Co2+, CN-, NO3-, SO42-, S2-, PO43-, Cl-, anion active surfactants, crude oils, Cl2 < 4 ppm
Komunikacija: Modbus via SC200 or SC1000 controller
Materiali v stiku z vodo: Probe body: CPVC, Polyurethane, Viton, Noryl, 316 Stainless Steel 1.4404 (AISI 316L)
Sensor cap: Acrylic
Material ohišja: Stainless
Merilno območje: 0.01 - 20.00 mg/L
  0 - 200 % air saturation
Možnosti: None
Natančnost: ± 0,05 ppm, < 1 ppm
  ± 0,2 ppm (> 5 ppm)
  ±0.1 mg/L @ DO < 5 mg/L
  Temperatura: ± 0,2 °C
Navoj senzorja: 1" NPT outer thread
Območje tlaka: Najv. 3,5 bar
Ocene nevarnih lokacij: ETL listed (cETLus marked) to Canadian and US General Safety and Hazardous (Class l, Div. 2) Locations
Oddaljenost (maksimalna) senzorja od kontrolne enote: 1000 m
Oddaljenost prenosa: 1000 m maximum when used with a junction box
  100 m maximum when used with an extension cable
Odzivni čas : T95 < 60 s
Odzivni čas - 20 °C : T90 < 40 s
pH območje: 0 - 12 pH
Pogoji skladiščenja: -20 °C - 70 °C
Ponovljivost: ±0.1 mg/L
Potopna globina: 15 m, 1050 kPa, maximum
Razred zaščite: IP68
Region: Global
Relativna vlažnost: 95 % (non-condensing)
Resolucija: 0.01 mg/L DO
Sensor cable: 10 m terminated with quick-disconnect plug
Stopnja pretoka: None required
Temperaturni senzor: PT100 integrated, external sensor
Teža: 1.0 kg (probe only)
Umertivena metoda: All probes are factory calibrated and ready to use
Air calibration: one point, 100% water saturated air
Sample calibration: comparison to standard instrument