Alarm:
1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analogni izhodi:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Avtomatsko čiščenje:
Da
Certificiranje:
CE compliant / ETL certified
Demineralised water:
For rinsing / dilution
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimenzija (v x š x g):
690 mm x 465 mm x 330 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancija:
24 mesecev
Ime metode:
Colorimetric
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interference :
Substances in amounts normally found in potable water will likely not interfere. Bromide, iodidie and cyanide register as equivalent chloride concentrations. Sulfide, thiosulfate and sulfite interfere but can be removed by treatment with hydrogen peroxide. Orthophosphate > 25 mg/L interferes by precipitating as silver phosphate. Iron > 10 mg/L interferes. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Kakovost vzorca:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Kiveta za vzorec:
10 min (dilution + 5 min)
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Meja zaznavanja:
1 mg/L
Merilna metoda:
Colorimetric measurement at 546 nm using diphenylcarbazide method, conform with standard method APHA 3500-Cr (B)
Merilno območje:
1 ‐ 10 mg/L
1 - 10 mg/L Cl-
Optional:
4 - 40 mg/L Cl- (with internal dilution)
8 - 80 mg/L Cl- (with internal dilution)
100 - 1000 mg/L Cl- (with internal dilution)
Moč:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Model:
EZ1005
Način premora:
100 - 240 VAC, 50/60 Hz
Odvod:
Atmospheric pressure, vented, min. Ø 32 mm
Parameter:
Chloride
Precision:
Better than 2% full scale range for standard test solutions
Pretok vzorca:
100 - 300 mL/min
Razred zaščite:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Reagent Requirements:
Keep between 10 - 30 °C
Št. vzorca:
1, 2, 4, or 8
Temperatura okolja:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura vzorca:
10 - 30 °C
Teža:
25 kg
Umeritev:
Automatic, 2-point; frequency freely programmable
Validation:
Automatic; frequency freely programmable
Vsebina paketa:
EZ1005 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 2.5L Reagent Container with Fittings (for Acid & Cleaning Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for AgNO₃ Solution)
Vzorec: tlak:
By external overflow vessel