CLT10 sc Total chlorine sensor with grab sample (on panel)

CLT10 sc Total chlorine sensor with grab sample (on panel)
Številka izdelka: LXV45B.98.21022
Cena na enoto Kontakt
Odprema v 2 tednih

Compatible with Hach's "Plug and Play" Digital Controllers

The CLF10 sc and CLT10 sc Chlorine Analysers are designed to be used with all Hach digital SC controllers. Set up is fast and easy, and the system is compatible with the full line of Hach digital sensors.

No Reagent Replacement, No Waste Stream

Chlorine measurement with the CLF10 sc and CLT10 sc amperometric technology eliminates the need for routine reagent replacement and waste stream management.

EPA Compliance

The CLF10 sc and CLT10 sc analysers can be used for reporting chlorine residual measurements in accordance with EPA Method 334.0. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required for Method 334.0.

Advanced Diagnostics

CLF10 sc and CLT10 sc advanced diagnostics alert you when the process has changed or the instument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting sample flow sensor for notification of insufficient sample flow.

Real-time Process Monitoring and Control

The CLF10 sc and CLT10 sc Chlorine Analysers allow for real-time monitoring and control of disinfection processes.

Vsebina paketa

CL10sc Panel with Grab Sample, 1 m Digital Extension Cable, Panel Manual, Chlorine Sensor Manual

Specifikacije

Certificiranje: CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller
Delovna temperatura: 0 - 45 °C
Dimenzija: Sensor Only: 195 mm x 25 mm (7.68 in x 0.98 in)
Dolžina: Sensor: 195 mm
Dolžina kabla: 1 m (between gateway and SC controller)
Garancija: 24 mesecev
Interference : Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits.
Total chlorine: chlorine dioxide, ozone and chalk deposits
Kontrolna enota: Panel only
Material: Panel: Corrosion-resistant, fully-submersible (stainless steel, PVC, silicon rubber and polycarbonate)
Meja zaznavanja: 30 ppb (0.03 ppm)
Merilna metoda: Reagentless, electrochemical, three-electrode amperometric system
Merilno območje: 0 - 20 ppm Cl₂
Možnosti: Panel Only
Način montaže: Panel or Wall Mounting
Napetost (V): None
Natančnost: Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
  Free Chlorine: ± of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
  Total Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
  Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5
Območje tlaka: 0.5 bar, no pressure impulses and/or vibrations
Odzivni čas : Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
  Total Chlorine: 100 s or less for 90% change (T90)
at a stable temperature and pH
Opis: CLT10sc Total Chlorine Sensor with Grab Sample (On Panel)
Parameter: Total Chlorine
pH Monitoring Required?: None
pH območje: 4 - 9 pH
Pogoji skladiščenja: -20 °C - 60 °C dry
Ponovljivost: 30 ppb ali 3%, kar je večje
Povezava s kablom: 5 pin, M12 connector
Pretok vzorca: 30 - 50 L/h, optimal is 40 L/hour
Priključek linije vzorca: ¼" OD
Priključek odvoda: ½" ID
Stopnja pretoka: Metric
Temperatura vzorca: 5 - 45 °C
Temperaturna kompenzacija: Internal temperature compensation in the sensor
  Internal temperature compensation in the sensor
Teža: Approximately 5.5 kg
Panel and empty panel-mounted components only
Umertivena metoda: 1-point or 2-point (zero and slope) calibration
Vsebina paketa: CL10sc Panel with Grab Sample, 1 m Digital Extension Cable, Panel Manual, Chlorine Sensor Manual
Width: 495.3 mm
Združljivost kontrolnih enot: SC200, SC1000, SC4500. All controllers sold separately